Recall Z-0299-2011
Ethicon Endo-Surgery Inc · initiated September 27, 2010
Product
REALIZE Adjustable Gastric Band-C Pak with Endoscopic Dissector, Calibration Tube and 15mm XCEL" trocar. Product Code RLZB32DGT, Lot #: G4RE81, G4RK6N, G4RL6Z, G4T824, G4T98X, G4TN3R, G4TP1A & G4TR4J.
Codes / lots: Product Code RLZB32DGT, Lot #: G4RE81, G4RK6N, G4RL6Z, G4T824, G4T98X, G4TN3R, G4TP1A & G4TR4J.
Reason for recall
The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate from its original position on the Locking Connector. This may result in: Tubing Strain Relief migrating on the Tubing; Kinking of the Tubing resulting in difficulty/inability to adjust the fluid volume within the Band; or a free foreign body in the abdominal wall or abdominal (peritoneal) ca
FDA-determined cause: Other
Action taken
Ethicon Endo-Surgery issued an Urgent : Devoced Recall Event 2240 letter dated October 8, 2010 to all consignees identifying the affected devices and actions to be taken. Consignees were instructed: DO NOTt use product from the affected lots, but implanted devices do not need to be removed as part of this recall. Examine inventory and remove the affected product. If they DO have the affected product, complete the Business Reply Form, keeping a copy for records, and place the original form in the box with the product to be returned, affix the pre-printed label, and return to Stericycle. If they DO NOT have the affected product, complete the Business Reply Form, and fax it to Stericycle. A redesigned replacement product will be sent out to consignees within 30-60 days of product return. Ethicon Endo-Surgery can be contacted at 513 337-8208 concerning this recall.
Details
- Recall number
- Z-0299-2011
- Initiated
- September 27, 2010
- Posted by FDA
- November 2, 2010
- Terminated
- November 18, 2011
- Product code
- LTI – Implant, Intragastric For Morbid Obesity
- Quantity
- 138 pieces
- Distribution
- Worldwide Distribution: USA including the following states: AK, AL, AR, AZ, CA, CO, CT, Washington DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, TN, TX, UT, VA, VT, WA, WI and WV, and the following countries: ARAB EMIRATES, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA…
- Recalling firm
- Ethicon Endo-Surgery Inc, Cincinnati, OH
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0289-2011 | REALIZE Adjustable Gastric Band, Injection Port and Applier Product Code: RLZB22…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0290-2011 | REALIZE Adjustable Gastric Band Pak with 15mm XCEL" trocar Product Code RLZB22T, Lot #:…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0291-2011 | REALIZE Adjustable Gastric Band Pak with Endoscopic Dissector Product Code RLZB22D1, Lot…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0292-2011 | REALIZE Adjustable Gastric Band Pak with Endoscopic Dissector and 15mm XCEL" trocar…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0293-2011 | REALIZE Adjustable Gastric Band Pak with Endoscopic Dissector and Calibration Tube…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0294-2011 | REALIZE Adjustable Gastric Band Pak with Endoscopic Dissector, Calibration Tube and 15mm…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0295-2011 | REALIZE Adjustable Gastric Band-C Product Code RLZB32, Lot #: ZJMBP7, ZJMBP8, ZJMBP9…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0296-2011 | REALIZE Adjustable Gastric Band-C Pak with 15mm XCEL" trocar Product Code RLZB32T, Lot…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0297-2011 | REALIZE Adjustable Gastric Band-C Pak with Endoscopic Dissector and 15mm XCEL" trocar…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0298-2011 | REALIZE Adjustable Gastric Band-C Pak with Endoscopic Dissector and calibration tube.…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0300-2011 | REALIZE Adjustable Gastric Band-C with Endoscopic Dissector. Product Code RLZB32D1, Lot…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0301-2011 | REALIZE Injection Port and Applier. Product Code RLZPT2, Lot #: ZHLBCG, ZHPBB8, ZJBBBD…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0302-2011 | SAGB QuickClose Swedish Adjustable Gastric Band including Velocity" Injection Port and…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0303-2011 | SAGB VC Swedish Adjustable Gastric Band including Velocity" Injection Port Applier…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |
| Z-0304-2011 | Velocity" Injection Port and Applier Product Code PT2XV, Lot #: 1110603, 10060505…The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate… | 09/27/2010Terminated |