Recall Z-0301-2020

Zimmer Biomet, Inc. · initiated October 10, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Coorad Morrey Total Elbow, Item Nos. 32810502501 32810502701 32810502900 32810503000 32810600012 32810600013 32855510796 32855510797 32855510799 32855510800 32855510801 32855510805 32855510822 32855510843 32855510844

Codes / lots: All product manufactured prior to January 2014.

Reason for recall

Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).

FDA-determined cause: Packaging change control

Action taken

The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time. 3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected. 4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components. 5. Complete the Certificate of Acknowledgement and return to the firm. 6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation. If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.

Details

Recall number
Z-0301-2020
Initiated
October 10, 2019
Terminated
May 19, 2020
Product code
JDC – Prosthesis, Elbow, Constrained, Cemented
Quantity
5,459,583 total devices
Distribution
Distributed nationwide.
Recalling firm
Zimmer Biomet, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0298-2020 Class IIBF Humeral Stem, Various Sizes, Item Nos. 00430000606 00430000613 00430000706…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0299-2020 Class IITrabecular Metal Reverse Shoulder System, Item Nos. 00434210613 00430900103 00434210613…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0300-2020 Class IISidus Stem-Free Shoulder Humeral Head, Item Nos. 0104555380 0104555400 0104555420…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0302-2020 Class IIVarious Custom Devices, Item Nos. 32855090246 32855090252 32855090256 32855090265…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0303-2020 Class IINCB Proximal Lateral Tibia, "NCB-PT LAX PROX 3H TIBIA", Various Sizes, Item Nos.…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0304-2020 Class IINCB PERIPROSTHETIC, "NCB PP PROX FEM PLATE", Various sizes, Item Nos. 0202263009…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0305-2020 Class IIZPLP Fibular Plate, Item Nos. 00234701504 00234701506 00234701508 00234701510…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0306-2020 Class IITM Ankle and Ankle Talus L/R, Item Nos. 00450001100 00450001200 00450001300 00450001400…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0307-2020 Class IIPeriarticular Plates - Tibial, Item Nos. 00234700304 00234700306 00234700308 00234700310…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0308-2020 Class IIPERIARTICULAR PLATES - FEMUR, Item Nos. 00234700106 00234700108 00234700110 00234700112…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0309-2020 Class IIProx Lat Tib Lock Plate and Dist Med Tib Lock Plate Item Nos. 47235700306 47235700308…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0310-2020 Class IIDIST LAT FEM LCK PLATE STERILE, Item Nos. 47235710106 47235710110 47235710114…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0311-2020 Class IISM DIS VOL/DORS RAD LAT COL PLATE, Item Nos. 47235801404 47235802102 47235802103…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0312-2020 Class IIUniversal Locking Plates Sterile, Item Nos. 47492800207 47492800303 47492800402…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0313-2020 Class IIPERIARTICULAR PLATES SHOULDER, PERI PROX LAT HUMERAL, Item Nos. 00234800104 00234800106…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0314-2020 Class IIPeriarticulare Plates - Elbow, Humeral and Ulna Plates, Item Nos. 00234800504…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0315-2020 Class IIPERI DIST VOLAR/DORSAL T/DORSAL DELTA RADIAL, Item Nos. 00234801303 00234801304…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0316-2020 Class IIPROX LAT HUMERAL LOCK PLATE, Item Nos. 47235800104 47235800106 47235800108 47235800112…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0317-2020 Class IIPeriarticular locking plate elbow-sterile, Item Nos. 47235800503 47235800509 47235800511…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0318-2020 Class IIVarious Hip Joint Replacement Systems, Item Nos. 00151504232 00151504432 00151504636…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0319-2020 Class IICLS Brevius Stem Kinectiv, hip prosthesis, various sizes. Item Nos. 0100296050…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0320-2020 Class IIFitmore Hip Stem, hip prosthesis, various sizes, Item Nos. 0100551102 0100551103…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0321-2020 Class IIVarious Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0322-2020 Class IIVarious Surgical Instruments: Screws and Reamers, Item Nos. 00579104100 00579104200…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0323-2020 Class IIVarious Devices for Limb salvage arthroplasty, Item Nos. 00585000011 00585001201…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0324-2020 Class IILPS-MOB ART SURF IMPL D 9MMNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0325-2020 Class IIVarious Knee Systems, Item Nos. 32855411838 32855411840 32855411842 32855411843…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0326-2020 Class IIVarious fixation systems and accessories, Item Nos. 00118100522 00118100527 00118100530…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0327-2020 Class IICABLE-READY SHORT GTR and CABLE-READY LONG GTR, Item Nos. 00223200201 00223200202Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated
Z-0328-2020 Class IIVarious fixation systems and accessories, Item Nos. 00224000916 00224000918 00224000920…Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). 10/10/2019Terminated