Recall Z-0313-2025

Beckman Coulter, Inc. · initiated September 30, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Codes / lots: UDI/DI 15099590231118, all lots

Reason for recall

The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.

FDA-determined cause: Process design

Action taken

Beckman Coulter issue and URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2024 via letter (USPS) and email. The notice explained the issue, the impact to the patient, and requested the use of alternative lot of Access Cortisol reagent. In the event the laboratory cannot generate a passing calibration curve use alternative methodology including alternative Beckman Coulter Access Immunoassay Systems. Beckman Coulter will release an updated Access Cortisol APF on the DxI 9000 Access Immunoassay Analyzer that includes the correct calibration curve acceptance parameter at the S1 calibrator level. A Beckman Coulter representative will contact you when the updated APF is available. The notice is to be shared with laboratory staff and forwarded to all parties to whom the the affected product was distributed.

Details

Recall number
Z-0313-2025
Initiated
September 30, 2024
Posted by FDA
November 6, 2024
Product code
CGR – Radioimmunoassay, Cortisol
Quantity
149 units
Distribution
Worldwide distribution.
Recalling firm
Beckman Coulter, Inc., Chaska, MN