Recall Z-0314-2017

Bbi Solutions Oem Ltd · initiated June 1, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood

Codes / lots: Lot Numbers containing test reagent lot # 1012-34437): BU183A, BU183B, BU184A, BU185A, BU185B Expiration Date: 31 JUL 2016

Reason for recall

A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.

FDA-determined cause: Nonconforming Material/Component

Action taken

BBI Solutions sent an Urgent Medical Device Recall letter dated June 30, 2016, to all affected customers. Customers were advised to immediately discontinue use of the product, consider the requirement for review of reported test results, complete and return the customer acknowledgement form to BBI Solutions and to complete and return the customer reconciliation form to the distributor. Customers who have forwarded product to other sites, should inform those customers of the product recall and provide them with a copy of the notice. Customers with questions were instructed to call Tel: +44(0)1382 564093.

Details

Recall number
Z-0314-2017
Initiated
June 1, 2016
Terminated
June 14, 2018
Product code
GHH – Fibrin Split Products
Quantity
420 units (183 units quarantined)
Distribution
US distribution to CT and IN
Recalling firm
Bbi Solutions Oem Ltd, Dundee