Recall Z-0315-2008

Beckman Coulter Inc · initiated April 18, 2007

Terminated

Product

COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer)

Codes / lots: Part Number 6605381. All software versions

Reason for recall

Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th

FDA-determined cause: Software design

Action taken

Product Corrective Action (PCA) letters were mailed on April 18,2007, to all GENS, LH500, LH750 and LH780 accounts. They are informed that a potential misidentification can occur in 2 scenarios. 1. When a positive identifier (Sample ID or CassetteIPosition) is manually edited to a positive identifier that is already in the ToDo list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2. When manually entering a Patient ID, if a blank space is entered between the characters in the Patient ID field, the system will only accept the characters before the space.----Short term fix----- The letter explains the issues and provides the customer work around. (Letters were sent by US mail). A customer response form was provided. The root cause of the problem is still under investigation, software upgrades, when approved will be made to devices.

Details

Recall number
Z-0315-2008
Initiated
April 18, 2007
Posted by FDA
December 22, 2007
Terminated
April 16, 2012
Product code
GKL – Counter, Cell, Automated (Particle Counter)
Quantity
815 units to the US; 95 units to Canada
Distribution
World wide: USA and Canada
Recalling firm
Beckman Coulter Inc, Brea, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0312-2008 COULTER LH750 Analyzer, Part Number 6605632, Beckman Coulter, Brea, CA. (Beckman…Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is… 04/18/2007Terminated
Z-0313-2008 COULTER LH780 Analyzer , Part Number: 723585, Beckman Coulter, Brea, CA. (Hematology…Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is… 04/18/2007Terminated
Z-0314-2008 COULTER LH 500 Series System, Part Number: 178833, Beckman Coulter, Brea, CA.…Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is… 04/18/2007Terminated