Recall Z-0316-2024

Atrium Medical Corporation · initiated September 18, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Ocean Water Seal Chest Drains (Ocean): Products REF Number Product Name (1)2002-000 DRAIN, OCEAN SINGLE W/AC,S; (2)2002-040 DRAIN, OCEAN SINGLE,PEDI CONNECTOR; (3)2002-100 DRAIN, OCEAN SINGLE W/AC; (4)2002-300 DRAIN, OCEAN SINGLE W/S; (5) 2002-400 DRAIN, OCEAN SINGLE ; (6)2012-320 DRAIN, OCEAN PEDI W/S; (7)2050-000 DRAIN, OCEAN BRU W/AC,S

Codes / lots: 1)2002-000 DRAIN, OCEAN SINGLE W/AC,S 20650862100017 (2)2002-040 DRAIN, OCEAN SINGLE,PEDI CONNECTOR 20650862100345 (3)2002-100 DRAIN, OCEAN SINGLE W/AC 20650862100093 (4)2002-300 DRAIN, OCEAN SINGLE W/S 20650862100109 (5) 2002-400 DRAIN, OCEAN SINGLE 20650862100215 (6)2012-320 DRAIN, OCEAN PEDI W/S 20650862101021 (7)2050-000 DRAIN, OCEAN BRU W/AC,S 20650862103018 All lots within labeled product expiry Manufacturing Dates: Any product manufactured on and after July 21, 2020 Distribution Dates: Any product shipped on and after August 6, 2020

Reason for recall

Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains

FDA-determined cause: Error in labeling

Action taken

UPDATE: April 2024. Getinge issued letter Getinge Thoracic Drainage Labeling Updates: New Instructions for Use Precaution for Oasis and Express Chest Drains to existing customers April 17, 2024 (US) and April 23, 2024 (OUS). As a result, inventory will be a mix of the current IFU and new IFU with the updated precaution statement until the current IFU-containing products are depleted. Reminder: In September 2023, Getinge initiated Medical Device Correction 3011175548-08/18/2023-001-C which requested your facility to post a copy of the Notice on Page 4 of the notification in all inventory locations where the devices are stored. Since the updated IFU now contains this new Precaution, the Page 4 Notice may be removed from all inventory locations once all affected product with expiry labeled March 7th, 2024 and earlier has been depleted. Please contact your Getinge Inside Sales Representative should you need an electronic copy of the new Instructions for Use. Atrium/Getinge issued URGENT MEDICAL DEVICE Correction letter on September 18, 2023 via FedEx 2 day delivery on or before, September 20, 2023. Letter states reason for recall, health risk and action to take: facility can continue use of the device. No devices need to be returned. " Please ensure that all Atrium Ocean, Oasis, or Express Single Collection Chest Drain users at your facility are aware of this Safety Notice and post a copy of the Notice on Page 4 in all inventory locations within your facility where the devices are stored Notification of the release of the updated IFU containing the new Precaution will be communicated to all customers upon release, including reminder the Notice on Page 4 may be removed upon the facility s receipt of product with the updated IFU. " Please forward this information to all current and potential Atrium Ocean, Oasis, or Express Single Collection Chest Drain users within your hospital / facility. " If you are a distributor who has shipped any affected products to custom

Details

Recall number
Z-0316-2024
Initiated
September 18, 2023
Posted by FDA
November 16, 2023
Product code
CAC – Apparatus, Autotransfusion
Quantity
31270 cases: 29942 cases US; 1328 cases OUS
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Brazil, Brunei Darussalam, Canada, Chile, China, Czech Republic Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Korea, Republic of Kuwait, Latvia, Lithuania, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Montenegro, Netherlands, New…
Recalling firm
Atrium Medical Corporation, Merrimack, NH

Official FDA recall record