Recall Z-0321-2026

Surgical Theater Inc · initiated September 16, 2025

Class II Completed

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Surgical Navigation Advanced Platform Patient Engagement (PE SNAP), software version SNP.3.12.0, imaging processing radiological system

Codes / lots: UDI/DI 00863157000374

Reason for recall

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

FDA-determined cause: Under Investigation by firm

Action taken

Surgical Theater issued an URGENT: Medical Device Software correction notice dated September 17, 2025 to its consignees on 09/22/2025 via email. The notice explained the issue, risk to health, and provided the following: Customers can continue to use the software version(s) noted above until a software patch is made available by Surgical Theater. Until the software patch is installed, please follow these steps: 1. Read and share this notice with all relevant staff. 2. Continue to use the 3D Fusion functions to review the fusion results for all case files to verify alignment of secondary layer(s) to primary layer and to adjust fusion as necessary, and 3. Acknowledge your receipt of this notification by completing and returning the included acknowledgement form via email to quality@surgicaltheater.com. A software patch is currently undergoing verification and should be available in the next 10 days. Surgical Theater has revised the 2D Image Fusion function to ensure the x, y and z axis rotation values are consistently applied when generating the aligned Axial, Sagittal and Coronal images. In addition, the rotate left/right buttons in all three planes were modified so the volume rotation stays consistent with the respective plane orientation. A Surgical Theater representative will contact your facility to schedule installation once the patch is available. For questions regarding this notification contact: Kevin Murrock, Sr. Director of Quality & Regulatory, Tel. 330-472-6520 (Monday through Friday, 9:00 AM to 5:00 PM, Eastern Time), email: kmurrock@surgicaltheater.com

Details

Recall number
Z-0321-2026
Initiated
September 16, 2025
Posted by FDA
October 24, 2025
Product code
LLZ – System, Image Processing, Radiological
Quantity
50 units
Distribution
Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Recalling firm
Surgical Theater Inc, Beachwood, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0319-2026 Class IISuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological…A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image… 09/16/2025Completed
Z-0320-2026 Class IISurgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging…A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image… 09/16/2025Completed
Z-0322-2026 Class IISyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software…A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image… 09/16/2025Completed
Z-0323-2026 Class IIEndo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software…A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image… 09/16/2025Completed
Z-0324-2026 Class IISpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software…A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image… 09/16/2025Completed
Z-0325-2026 Class IISurgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented realityA software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image… 09/16/2025Completed