Recall Z-0327-2012

Varian Medical Systems, Inc. · initiated September 30, 2008

Terminated

Product

GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 200 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany. The interstitial plastic needles with 2 mm diameter are designed for interstitial treatments in areas such as head and neck, gynecology, breast or prostate. In order to avoid bending and to facilitate the introduction of the needles, each needle is supplied with a mandrin.

Codes / lots: Part number GM11007570

Reason for recall

Plastic needle tip could possibly crack or separate.

FDA-determined cause: Nonconforming Material/Component

Action taken

Varian Medical Systems sent a "CUSTOMER TECHNICAL BULLETIN" letter dated September 30, 2008 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed users to inspect needle tips before and after every clinical application to prevent needle tip separation or cracking.

Details

Recall number
Z-0327-2012
Initiated
September 30, 2008
Posted by FDA
December 1, 2011
Terminated
December 7, 2011
Product code
JAQ – System, Applicator, Radionuclide, Remote-Controlled
Quantity
1716 parts
Distribution
Worldwide Distribution - USA (nationwide) including the states of Arizonia, California, Florida, Illinois, Indiana, Michigan, North Carolina, and Vermont, and the countries of Argentina, Austria, Belgium, Canada, Estonia, France, Germany, India, Norway, Pakistan, Russia, Slovenia, So. Africia, Spain, Switzerland, United Kingdom, and Ukraine.
Recalling firm
Varian Medical Systems, Inc., Charlottesville, VA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0326-2012 GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 113 mm. Remote controlled…Plastic needle tip could possibly crack or separate. 09/30/2008Terminated
Z-0328-2012 GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 320 mm. Remote controlled…Plastic needle tip could possibly crack or separate. 09/30/2008Terminated