Recall Z-0330-2012

Varian Medical Systems, Inc. Oncology Systems · initiated October 6, 2011

Terminated

Product

Varian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm diameter, GammaMed 12i; Ref: BT-00264, Model 2-H63; Product is distributed by Varian Medical Systems Inc., Palo Alto, CA and Manufactured by Varian Medical Systems Haan GmbH, Bergische Strasse 16, D-42781 Haan, Germany. The Varian Medical System GammaMed Afterloader system is a remotely controlled intelligent robotic device intended to be used to deliver high dose-rate radiation therapy, in close proximity to and from within diseased tissues. Palliative care and treatments with curative intent to both benign and non-benign disease using interstitial, intracavitary, intraluminal, intra and inter-operative brachytherapy or combinations of any these treatment types are possible using this device.

Codes / lots: CODES: H630164 ,H630703 ,H630727 ,H630756 ,H630787 ,H630810 ,H630829 ,H630850 ,H630868 ,H630888 ,H630165 ,H630705 ,H630728 ,H630757 ,H630789 ,H630811 ,H630832 ,H630851 ,H630869 ,H630889 ,H630168 ,H630706 ,H630729 ,H630758 ,H630790 ,H630812 ,H630833 ,H630852 ,H630870 ,H630891 ,H630170 ,H630711 ,H630732 ,H630760 ,H630792 ,H630813 ,H630834 ,H630854 ,H630871 ,H630892 ,H630205 ,H630712 ,H630733 ,H630764 ,H630793 ,H630814 ,H630835 ,H630855 ,H630872 ,H630895 ,H630206 ,H630713 ,H630736 ,H630769 ,H630794 ,H630815 ,H630837 ,H630857 ,H630877 ,H630896 ,H630210 ,H630714 ,H630737 ,H630771 ,H630795 ,H630817 ,H630839 ,H630858 ,H630878 ,H630897 ,H630211 ,H630716 ,H630739 ,H630774 ,H630798 ,H630818 ,H630841 ,H630859 ,H630879 ,H630898 ,H630214 ,H630718 ,H630741 ,H630776 ,H630801 ,H630820 ,H630842 ,H630860 ,H630880 ,H630901 ,H630215 ,H630719 ,H630744 ,H630777 ,H630802 ,H630822 ,H630843 ,H630863 ,H630881 ,H630902 ,H630216 ,H630720 ,H630747 ,H630778 ,H630804 ,H630825 ,H630845 ,H630864 ,H630884 ,H630903 ,H630219 ,H630721 ,H630749 ,H630781 ,H630807 ,H630826 ,H630847 ,H630866 ,H630885 ,H630904 ,H630701 ,H630722 ,H630753 ,H630784 ,H630809 ,H630828 ,H630848 ,H630867 ,H630886

Reason for recall

The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector.

FDA-determined cause: Device Design

Action taken

Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION LETTER/FIELD SAFETY NOTICE" letter dated October 6, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Contact the Varian Help Desk at 1-888-827-4265 for questions regarding this notice.

Details

Recall number
Z-0330-2012
Initiated
October 6, 2011
Posted by FDA
December 1, 2011
Terminated
October 22, 2012
Product code
JAQ – System, Applicator, Radionuclide, Remote-Controlled
Distribution
Worldwide Distribution--USA (nationwide).
Recalling firm
Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0329-2012 Varian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm…The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the… 10/06/2011Terminated