Recall Z-0332-2009
Boston Scientific Corporation · initiated October 7, 2008
Product
Boston Scientific Easy Core" Biopsy System, 18 ga. x 21 cm, REF 43-454, UPN M001434541, Sterilized with ethylene oxide gas. UPN for inner pouch M001434540. Indicated for use endoscopically or percutaneously to retrieve tissue sampling of soft organs, tumors or masses for histological analysis. Soft tissue sampling includes, but is not limited to, organs such as breast, liver, kidney or prostate.
Codes / lots: 11520676 and 11551300
Reason for recall
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
FDA-determined cause: Device Design
Action taken
Consignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated October 21, 2008. The letter was addressed to The Hospital Field Action Contact. The letter described the problem and products being recalled. The letter stated that it was very important that the attached product table be used carefully and consider both inner and outer packaging product codes when searching for affected/recalled product as the UPN numbers on the inner and outer labeling are different. Requested to discontinue use of product, identify, and segregate and return affected products to BSC. Letter also requested to complete and return the Reply Verification Tracking Form.
Details
- Recall number
- Z-0332-2009
- Initiated
- October 7, 2008
- Posted by FDA
- November 28, 2008
- Terminated
- December 10, 2011
- Product code
- FCG – Biopsy Needle
- Quantity
- 3808 total for all products
- Distribution
- Nationwide Distribution --- including states of AL, AZ, AR, CA, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO,MT, NE , NV, NH, NH, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and PR.
- Recalling firm
- Boston Scientific Corporation, Maple Grove, MN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0328-2009 | Boston Scientific Easy Core" Biopsy System, 18 ga. x 10 cm, REF 43-452, UPN MOO1434521…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0329-2009 | Boston Scientific Easy Core" Biopsy System, 15 ga. x 10 cm, REF 43-450, UPN M001434501…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0330-2009 | Boston Scientific Easy Core" Biopsy System, 15 ga. x 15 cm, REF 43-451, UPN M001434511…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0331-2009 | Boston Scientific Easy Core" Biopsy System, 18 ga. x 15 cm, REF 43-453 , M001434531…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0333-2009 | Boston Scientific Easy Core" Biopsy System, 15 ga. x 10 cm, REF 43-456, UPN M001434561…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0334-2009 | Boston Scientific Easy Core" Biopsy System, 15 ga. x 15 cm, REF 43-457, UPN M001434571…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0335-2009 | Boston Scientific Easy Core" Biopsy System, 18 ga. x 10 cm, REF 43-458, UPN M001434581…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0336-2009 | Boston Scientific Easy Core" Biopsy System, 18 ga. x 15 cm, REF 43-459, UPN M001434591…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |
| Z-0337-2009 | Boston Scientific Easy Core" Biopsy System, 18 ga. x 21 cm, REF 43-460, UPN M001434601…Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking… | 10/07/2008Terminated |