Recall Z-0338-2022

Covidien Llc · initiated November 4, 2021

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

Puritan Bennett 980 Series Ventilator, 980A1ENNISB

Codes / lots: GTIN/UDI - 10884521201897 Serial Numbers: 35B1700507, 35B1700533, 35B1700539, 35B1700465, 35B1700580, 35B1700530, 35B17005363, 5B1700542, 35B1700509, 35B1701614, 35B1700527, 35B1700537, 35B1700541, 35B1700508 & 35B1701569.

Reason for recall

Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

FDA-determined cause: Reprocessing Controls

Action taken

On November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use. Actions you should take Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies. Notify all personnel in all care environments in which the affected PB980 series ventilators are used about this medical device correction. If your facility has distributed affected PB980 series ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. Complete the attached form and return it as directed to confirm your receipt and understanding of this information. If you are aware of any incidents related to this issue, please contact our Technical Support Department immediately at 1-800-255-6774, option 4, and then option 1 to provide information regarding those events so regulatory reporting obligations can be fulfilled. Work with the Medtronic Technical Support Department if you require assistance finding alternative ventilation devices. Beginning November 5, 2021, Medtronic Technical Service will contact all affected customers to schedule a service appointment for all potentially affected ventilators. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.

Details

Recall number
Z-0338-2022
Initiated
November 4, 2021
Posted by FDA
December 9, 2021
Terminated
July 9, 2024
Product code
CBK – Ventilator, Continuous, Facility Use
Quantity
278 total
Distribution
Worldwide distribution - US Nationwide.
Recalling firm
Covidien Llc, Mansfield, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0339-2022 Class IPuritan Bennett 980 Series Ventilator, 980A3ENNISBManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0340-2022 Class IPuritan Bennett 980 Series Ventilator, 980S1ENDICUManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0341-2022 Class IPuritan Bennett 980 Series Ventilator, 980S3ENDICUManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0342-2022 Class IPuritan Bennett 980 Series Ventilator, 980U1ENASAAManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0343-2022 Class IPuritan Bennett 980 Series Ventilator, 980U1ESDIUCManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0344-2022 Class IPuritan Bennett 980 Series Ventilator, 980U3ESDIUUManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0345-2022 Class IPuritan Bennett 980 Series Ventilator,980X1ENAIQCManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0346-2022 Class IPuritan Bennett 980 Series Ventilator, 980X1ENDIPCManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0347-2022 Class IPuritan Bennett 980 Series Ventilator, 980X1ENDIUUManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0348-2022 Class IPuritan Bennett 980 Series Ventilator, 980X1ENNISBManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0349-2022 Class IPuritan Bennett 980 Series Ventilator, 980X1JADIJJManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0350-2022 Class IPuritan Bennett 980 Series Ventilator, 980X1PLDIPCManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0351-2022 Class IPuritan Bennett 980 Series Ventilator, 980X2ESDIECManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0352-2022 Class IPuritan Bennett 980 Series Ventilator, 980X3ENDIUUManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated
Z-0353-2022 Class IPuritan Bennett 980 Series Ventilator, 980X3JADIJJManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may… 11/04/2021Terminated