Recall Z-0342-2018

Spineart SA · initiated November 27, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

TRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.5 MM, LENGTH 18 MM, REF TRY-PS-45 18-S, Qty 2 TRYPTIK(R) 2C-Plate Anterior Cervical Plate System is intended to be used for temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumor, spondylolisthesis, spinal stenosis, deformity (i.e., scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions.

Codes / lots: batch: 3-5939

Reason for recall

Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the sketch/drawing of a self-drilling screw.

FDA-determined cause: Error in labeling

Action taken

Spineart sent an Urgent Medical Device letter by email dated November 27, 2017. The letter stated the following: "Out of an abundance of caution", we have decided to recall the concerned products. Please carefully follow instructions detailed below: 1. Immediately examine your inventory and quarantine product subject to recall. 2. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall by using this document. 3. Collect and quarantine all products. 4. Send back all products and the enclosed Response Form to Spineart Geneva E-mail: regulatory@spineart.com." For further questions, please call (949) 313-6400..

Details

Recall number
Z-0342-2018
Initiated
November 27, 2017
Terminated
August 27, 2019
Product code
KWQ – Appliance, Fixation, Spinal Intervertebral Body
Quantity
8 units
Distribution
US Distribution to the state of : CA
Recalling firm
Spineart SA, Geneva

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0336-2018 Class IITRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.0 MM, LENGTH 12 MM, REF TRY-PS-40…Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the… 11/27/2017Terminated
Z-0337-2018 Class IITRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.0 MM, LENGTH 14 MM, REF TRY-PS-40…Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the… 11/27/2017Terminated
Z-0338-2018 Class IITRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.0 MM, LENGTH 16 MM, REF TRY-PS-40…Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the… 11/27/2017Terminated
Z-0339-2018 Class IITRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.5 MM, LENGTH 12 MM, REF TRY-PS-45…Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the… 11/27/2017Terminated
Z-0340-2018 Class IITRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.5 MM, LENGTH 14 MM, REF TRY-PS-45…Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the… 11/27/2017Terminated
Z-0341-2018 Class IITRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.5 MM, LENGTH 16 MM, REF TRY-PS-45…Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the… 11/27/2017Terminated