Recall Z-0344-2014

Siemens Healthcare Diagnostics · initiated September 3, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ADVIA 1650 System Software: 1) V4.01, 2) V3.52. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.

Codes / lots: Product Codes: 1) 10318424, 2) 10285281

Reason for recall

Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems.

FDA-determined cause: Software design

Action taken

On September 3, 2013, Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice for the ADVIA 1200, 1650, 1800 and 2400 Chemistry Systems due to a software issue. The software issue has caused the Chemistry Systems to generate reports without error flags. Notices were sent to domestic customers via U.S. and outside the US the distribution is determined at the country level. The customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue at 877-229-3711 (prompt 18).

Details

Recall number
Z-0344-2014
Initiated
September 3, 2013
Posted by FDA
November 18, 2013
Terminated
March 7, 2017
Product code
JGS – Electrode, Ion Specific, Sodium
Quantity
892 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Vietnam.
Recalling firm
Siemens Healthcare Diagnostics, Tarrytown, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0343-2014 Class IIADVIA 1200 System Software: 1) V2.00, 2) V2.01. Performs assays for general and…Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging… 09/03/2013Terminated
Z-0345-2014 Class IIADVIA 1800 System Software V2.01. Performs assays for general and specialty chemistries…Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging… 09/03/2013Terminated
Z-0346-2014 Class IIADVIA 2400 System Software V4.01. Performs assays for general and specialty chemistries…Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging… 09/03/2013Terminated