Recall Z-0346-2024
Aizu Olympus Co., Ltd. · initiated September 25, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.
Codes / lots: Model Number (UDI-DI): TJF-145 (04953170339950), TJF-150 (04953170307553), TJF-Q180V (04953170229503), TJF-Q190V (04953170405563).
Reason for recall
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
FDA-determined cause: Under Investigation by firm
Action taken
Olympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Olympus is in the process of implementing a validated drying process at our repair centers. In the meantime, Olympus has established interim processes to perform an additional inspection of your scope for water prior to return. 1. Ensure your users are knowledgeable on the content of this letter. 2. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at olympus2806@sedgwick.com or by fax at 844-294-7184. For any questions about the acknowledgement form, please call the Sedgwick team at 888-667-1504. 3. Olympus requests that you report complaints, including infections, and endoscopes returned wet from our repair centers to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you require additional information, please do not hesitate to contact me at (647) 999-3203 or Cynthia.Ow@Olympus.com.
Details
- Recall number
- Z-0346-2024
- Initiated
- September 25, 2023
- Posted by FDA
- November 17, 2023
- Product code
- FET – Endoscopic Video Imaging System/Component, Gastroenterology-Urology
- Quantity
- 0 units
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- Aizu Olympus Co., Ltd., Aizuwakamatsu, 3 Chome
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0327-2024 Class II | Bronchoscope-for endoscopy and endoscopic surgery within the airways and…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0328-2024 Class II | Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0329-2024 Class II | Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0330-2024 Class III | Choleodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0331-2024 Class II | Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0332-2024 Class II | Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0333-2024 Class II | Gastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0334-2024 Class II | Gastrointestinal videoscope-indicated for use within the upper digestive tract…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0335-2024 Class II | Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0336-2024 Class II | Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0337-2024 Class II | Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0338-2024 Class II | Tracheal fiberscope-airway management, which includes observation to access airway…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0339-2024 Class II | Tracheal videoscope- For airway management, which includes observation to access airway…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0340-2024 Class II | Mobile airway scope-diagnosis and observation to access airway anatomy…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0341-2024 Class II | Colonoscope-indicated for use within the lower digestive tract (including the anus…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0342-2024 Class II | Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0343-2024 Class II | Small intestinal videoscope-For endoscopy and endoscopic surgery within the upper…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0344-2024 Class II | Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0345-2024 Class II | Ultrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0347-2024 Class II | Ureteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |
| Z-0348-2024 Class II | Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and…Channel air drying process was not validated, and that a small percentage of scopes returned to customers… | 09/25/2023Open, Classified |