Recall Z-0357-2020

TriMed Inc. · initiated September 28, 2018

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

TriMed Hex Driver 1.2, 55mm AO Cannulated, REF HXDRIVR-1.2C AO, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.

Codes / lots: Lot numbers 91500604 and 91600017

Reason for recall

The epoxy on Apex manufactured epoxy-coated instruments can pop off.

FDA-determined cause: Nonconforming Material/Component

Action taken

The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.

Details

Recall number
Z-0357-2020
Initiated
September 28, 2018
Product code
LXH – Orthopedic Manual Surgical Instrument
Quantity
467 hex drivers
Distribution
Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.
Recalling firm
TriMed Inc., Santa Clarita, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0355-2020 Class IITriMed Countersink 1.7mm, HCS, REF HSINK-1.7, nonsterile, Rx. The firm name on the label…The epoxy on Apex manufactured epoxy-coated instruments can pop off. 09/28/2018Open, Classified
Z-0356-2020 Class IITriMed Locking Drill Guide, 2.3mm, REF GUIDELCBS-2.3, nonsterile, Rx. The firm name on…The epoxy on Apex manufactured epoxy-coated instruments can pop off. 09/28/2018Open, Classified
Z-0358-2020 Class IITriMed Peg Guide Extender, REF PEG-XTNDR, nonsterile, Rx. The firm name on the label is…The spring on the PEG-Extender can rust due to the type of steel used to manufacture it. 09/28/2018Open, Classified
Z-0359-2020 Class IITriMed Wrist Hook Dorsal Impactor, REF IMPCT-WHD; and TriMed Wrist Hook Volar Impactor…The WHD/WHV impactors can rust over time due to an improper manufacturing process. 09/28/2018Open, Classified