Recall Z-0360-2014

Salter Labs · initiated November 1, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Salter Labs EtCO2 Sample line 15' (4.6 M) tube, male luer connectors. Rx only; Made in Mexico; Manufacturer:Salter Labs, 100 West Sycamore Road, Arvin, CA 93203. CO2 sample line.

Codes / lots: Model number: 4515-15-25; lot numbers 071513, 071513A and 120312.

Reason for recall

It has been determined that three lots of model number 4515-15-25 CO2 Sample Line 15' tube, male Luer lock connectors may contain a 30' tube instead of the 15' length.

FDA-determined cause: Labeling Change Control

Action taken

Salter Labs called the five customers that received product from the three lots, then follow-up immediately by sending a recall notification letter directly to each customer. Each letter will identify all shipments of subject product that they received. All customers letters were sent certified mail, return receipt requested.

Details

Recall number
Z-0360-2014
Initiated
November 1, 2013
Posted by FDA
November 19, 2013
Terminated
December 17, 2013
Product code
CCK – Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Quantity
91 cases (25 units per case)
Distribution
Distributed in the states of MS, TX FL,, WI, MI, VA, NY, MO, NC, and MS.
Recalling firm
Salter Labs, Arvin, CA