Recall Z-0361-2016

Quest Medical, Inc. · initiated October 28, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Model 5001102 MPS Delivery Set w/ arrest agent and additive cassettes, heat exchanger &10 ft. delivery tubing; Model 5001102-AS MPS Delivery Set with 6 ft. delivery tubing; Model 7001102 MPS Low Volume Delivery Set for use with cardiopulmonary bypass equipment.

Codes / lots: 0491795E04, 0492175E04, 0492185E06, 0492195E06, 0492615E06, 0492625E08, 0492635E08, 0489905Y04, 0490245Y06, 0493075U02, 0493085U02, 0493435U04, 0493675U04, 0493685U07, 494465U07, 0494475U09, 0494855G01, 0495215U11, 0495225U11, 0495505G02, 495515G02, 0496535G08, 0496875G08, 496885G08, 0497245S02, 0495205U11, 497255S04, 0497655S04, 0498535S06, 0498545S06, 0492205E03, 0493695U04, 0494485U07, 0495525G02, 0497265S02, 0495965G04, 0492765E07

Reason for recall

The products have been found to intermittently exhibit a seal failure during use.

FDA-determined cause: Equipment maintenance

Action taken

The recalling firm issued a press release on 10/28/2015. The press release was posted on FDA website on 10/29/2015. The recalling firm emailed affected consignees on 10/28/15 and sent out recall letters via USPS beginning 10/30/15.

Details

Recall number
Z-0361-2016
Initiated
October 28, 2015
Posted by FDA
December 2, 2015
Terminated
November 17, 2016
Product code
DTR – Heat-Exchanger, Cardiopulmonary Bypass
Quantity
18,761 units
Distribution
Distributed US (nationwide) and the countries of Canada, Japan, Saudi Arabia, Kuwait, South Africa, Switzerland, France, and Italy.
Recalling firm
Quest Medical, Inc., Allen, TX

Official FDA recall record