Recall Z-0363-2023

Randox Laboratories Ltd. · initiated November 1, 2022

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

IgM Immunoturbidimetric Assay-In vitro determination of IgA and IgM in serum and plasma Catalog Number: IM3834

Codes / lots: GTIN: 05055273204001 All lots

Reason for recall

Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to an elevation in Quality Control and Patient results of up to +13% for lgA and +51% for lgM respectively, which may lead to a delay in running patient samples or erroneous elevated test results

FDA-determined cause: Under Investigation by firm

Action taken

Randox Laboratories extended recall notification to the US Distribution center on 11/01/22 via email. The US distributor contacted customers on 11/1/22 via email . Letter states reason for recall, health risk and action to take: Review your instrument testing order in line with the Carryover Avoidance Technical Bulletin (RXTB-0136} " Update the RX user manual with the updated Carryover Avoidance Technical Bulletin (RXTB-0136} and ensure all operators are aware of the recommendations. Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. If you have any questions or concerns, please contact Randox Technical Services

Details

Recall number
Z-0363-2023
Initiated
November 1, 2022
Posted by FDA
December 3, 2022
Terminated
June 18, 2024
Product code
DFT – Igm, Antigen, Antiserum, Control
Quantity
20 kits US in total
Distribution
US Nationwide Distribution: CA, NC, NJ and WV,
Recalling firm
Randox Laboratories Ltd., Crumlin, North

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0362-2023 Class IIIgA Immunoturbidimetric Assay- In vitro determination of IgA and IgM in serum and plasma…Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run… 11/01/2022Terminated