Recall Z-0374-2026

Integra LifeSciences Corp. (NeuroSciences) · initiated September 26, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

CODMAN Disposable Perforator 9mm. Cranial Perforator.

Codes / lots: Model No. 261223; UDI: 10381780513605; Lot No. 5828553, 6264478, 6264481, 6264482, 7230672, 7230673, 7230674, 7230675, 7261971, 7268008, 7268009, 7268010, 7268011, 7268012, 7268013, 7268014, 7268015, 7268016, 7268017, 7268018, 7268019, 7282954, 7282955, 7293422, 7293423, 7293424, 7293425, 7293426, 7293427, 7293428, 7300667, 7300668, 7300669, 7300670, 7300672, 7300673, 7300674, 7371082, 7371083, 7379652, 7379922, 7379923, 7379924, 7379926, 7379928, 7379660, 7379925, 7379927, 7450559, 7450560, 7450561, 7450562, 7450563, 7450564, 7450565, 7450566, 7451273, 7451274, 7451275, 7451276, 7451277, 7451278, 7451279; Expiration Date: 2024-03-31 to 2029-07-31.

Reason for recall

Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.

FDA-determined cause: Process change control

Action taken

An URGENT: MEDICAL DEVICE RECALL notification dated 9/26/25 was mailed to consignees. Customers with affected devices are to be immediately removed and quarantined. Customers are to send their completed Acknowledgement Form to FCA3@integralife.com or by fax at 1-609-750-4220. Forward the recall notification to those who use the product for their awareness. Integra Customer Service will reach out to consignees to facilitate product return. Distributors and Sales Representatives with affected devices are to remove product from further distribution. They are to send their completed Acknowledgement Form to FCA3@integralife.com or by fax at 1-609-750-4220. Distributors are to create an acknowledgement form from the distributor to their customers and provide them with a copy of the recall notification. Sales Representatives are to receive further directions from Integra Customer Service after their Acknowledgement Form has been received. Consignees with any questions are to call 1-800-654-2873 or email custsvcnj@integralife.com.

Details

Recall number
Z-0374-2026
Initiated
September 26, 2025
Posted by FDA
November 3, 2025
Product code
HBF – Drills, Burrs, Trephines & Accessories (Compound, Powered)
Quantity
5,109 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czechia, Ecuador, France, French Guiana, Germany, Greece, India, Ireland, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Morrocco, Netherlands, New Zealand, Palestine, State of, Philippines, Portugal, Puerto Rico, Qatar, Republic of…
Recalling firm
Integra LifeSciences Corp. (NeuroSciences), Princeton, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0373-2026 Class ICODMAN Disposable Perforator 11mm. Cranial Perforator.Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm… 09/26/2025Open, Classified