Recall Z-0379-2017

Abbott Laboratories · initiated October 11, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.

Codes / lots: Lot: 31497UN14 Exp.: 17 Oct. 2016; Lot: 09353UNI15 Exp.: 12 Feb. 2017; Lot: 45216UN15 12 May 2017; Lot: 14596UN15 Exp.: 30 Jul. 2017; Lot: 37055UN15 Exp.: 12 Oct. 2017; Lot: 14782UN16 Exp.: 21 Apr. 2018

Reason for recall

Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).

FDA-determined cause: Under Investigation by firm

Action taken

Abbott Laboratories sent an Product Correction Immediate Action Required letter dated October 10, 2016, to all affected customers. Please review this letter with your Medical Director Taking the above information into consideration, you may continue to use ARCHITECT Lactic Acid reagent (LN 9D89-21) Please complete and return the Customer Reply Form If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter Please retain this letter for your laboratory records. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service. For further questions please call (847) 937-2550.

Details

Recall number
Z-0379-2017
Initiated
October 11, 2016
Posted by FDA
November 9, 2016
Terminated
May 17, 2017
Product code
KHP – Acid, Lactic, Enzymatic Method
Quantity
26, 837 kits
Distribution
Nationwide and Worldwide Distribution.
Recalling firm
Abbott Laboratories, Abbott Park, IL

Official FDA recall record