Recall Z-0391-2017

Trumpf Medical Systems, Inc. · initiated October 25, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

OR Table, Top U26 H V, Model # 1909793; OR Table Top U26 H V U, Model # 1909794; OR Table Top U24 H V, Model # 1909796; OR Table Top U24 H V U, Model # 1909797; OR Table Top U24 H V W, Model # 1909798; OR Table Top U14 H V, Model # 1909799 Product Usage: Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Patient transport on the operating table top from a patient transfer system to the operating theater or from the operating theater to a patient transfer system.

Codes / lots: Model # 190973 - Serial Number Range 102394262 - 102974610; Model # 1909794 - 102884276; Model # 1909796 - 102440024 - 102981676; Model # 1909797 - 102757564 - 102958935; Model # 1909798 - 102831498 - 102831519; Model # 1909799 - 102393683 - 102969656

Reason for recall

Trumpf has identified six incidents of pinched cables.

FDA-determined cause: Process control

Action taken

Trumpf Medical sent an Urgent Medical Device Correction letter dated October 7, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken The letter indicated that a authorized service provider will inspect the table tops and make necessary corrections. A confirmation notice was included with the letter which is to be returned. For contact Trupmpf Medical Systems, Inc. at 888-474-9359.

Details

Recall number
Z-0391-2017
Initiated
October 25, 2016
Posted by FDA
November 10, 2016
Terminated
February 6, 2017
Product code
JEA – Table, Surgical With Orthopedic Accessories, Ac-Powered
Quantity
4 (US) & 229 (Foreign)
Distribution
Worldwide Distribution - US Nationwide in the states of AZ and SC and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Iceland, Italy, Norway, Qatar, Slovenia, Spain, Sweden, Switzerland
Recalling firm
Trumpf Medical Systems, Inc., Charleston, SC