Recall Z-0408-2026
Greiner Bio-One North America, Inc. · initiated August 8, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322
Codes / lots: UDI-DI: 29120017574488(Rack), 39120017574485(Case). Lot: B250139Q, Expiration: 01/09/2026
Reason for recall
Blood collection tubes were manufactured with the label applied too high on the tube, which shifted the printed fill level mark upward, which could result in unnecessary sample rejection and repeat collections.
FDA-determined cause: Under Investigation by firm
Action taken
Greiner issued an Urgent Product Recall notice to its consignees on 08/19/2025 via email and fax. The notice explained the issue and requested the following: "Immediate Actions Required 1. Stop Distribution & Isolate Inventory Immediately cease distribution of the affected lot and segregate all remaining units. 2. Select ONE of the following notification options: A. Customer Notification by Greiner Bio-One o Complete the attached Product Disposition Site Confirmation form. o Email the form to recalls.us@gbo.com or fax to 877-217-7927. o Provide an Excel list of all customers including, if applicable, subdistributors who received the affected product, containing: . Facility name and address (city, state, ZIP) . Contact name, email, and phone number for Clinical/Laboratory Director or Supply Chain . Date(s) and quantity shipped o Greiner will handle direct customer notification and coordinate product return/credit. B. Customer Notification by Distributor o Use the customer letter provided by Greiner Bio-One. o Contact all customers who received the affected lot, including any sub-distributors. o Conduct follow-up with non-responders and document effectiveness checks per FDA guidance: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/industryguidance-recalls" For questions regarding this recall, please contact: Greiner Bio-One Technical Service, Phone: 800-515-8112, Fax: 877-217-7927, Email: recalls.us@gbo.com
Details
- Recall number
- Z-0408-2026
- Initiated
- August 8, 2025
- Posted by FDA
- October 28, 2025
- Product code
- GIM – Tubes, Vacuum Sample, With Anticoagulant
- Quantity
- 504,000
- Distribution
- US Nationwide distribution.
- Recalling firm
- Greiner Bio-One North America, Inc., Monroe, NC