Recall Z-0410-2026

Medistim Asa · initiated September 26, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No

Codes / lots: Lot Code: Serial number; Model number 5042; MQC04001 5043; MQC04001 5044; MQC14001 5061; MQU14011 5062; MQU14011 5064; MQU14011 5065; MQU14011 5066; MQU14011 5067; MQC04001 5037; MQU14011 5073; MQV12011 5038; MQU14011 5045; MQU14011 5063; MQU14011

Reason for recall

Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.

FDA-determined cause: Process control

Action taken

Firm notified impacted consignees by email on September 26, 2025. Consignees were informed that qualified service personnel will perform the correction of the device according to the standard service procedure.

Details

Recall number
Z-0410-2026
Initiated
September 26, 2025
Posted by FDA
October 29, 2025
Product code
DPW – Flowmeter, Blood, Cardiovascular
Quantity
14
Distribution
Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia.
Recalling firm
Medistim Asa, Oslo

Official FDA recall record