Recall Z-0412-2020

Randox Laboratories, Limited · initiated October 4, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Randox Liquid Cardiac Controls Catalogue Number CQ5051

Codes / lots: Catalogue Number: CQ5051 GTIN: 05055273207446 Batch/lot: 4243CK (Exp. 28Nov2019), 4246CK (Exp. 28Nov2019), 4249CK (Exp. 28Nov2019), 4260CK (Exp. 28Nov2019), 4311CK (Exp. 28May2020), 4317CK (Exp. 28May2020)

Reason for recall

Removal of claims for N-proBNP in the Liquid Cardiac Control model numbers CQ5051 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.

FDA-determined cause: Under Investigation by firm

Action taken

On October 1, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Inspect your stock and quarantine affected stock. - Replace the value sheet in the kit with the revised value sheet provided. - Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. - Discuss the contents of this notice with your Medical Director. - Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.

Details

Recall number
Z-0412-2020
Initiated
October 4, 2019
Terminated
July 6, 2020
Product code
JJY – Multi-Analyte Controls, All Kinds (Assayed)
Quantity
6,208
Distribution
US Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,
Recalling firm
Randox Laboratories, Limited, Crumlin

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0413-2020 Class IIRandox Liquid Cardiac Controls Catalogue Number CQ5052.Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5052 due to an observed… 10/04/2019Terminated
Z-0414-2020 Class IIRandox Liquid Cardiac Controls Catalogue Numbers CQ5053.Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed… 10/04/2019Terminated