Recall Z-0415-2009

Biomet Microfixation, Inc. · initiated October 24, 2008

Terminated

Product

Fredrick's Converse Retractor #5 F/O, Model Number: 01-0485. A fiber optic retractor used in plastic surgery procedures to retract tissue.

Codes / lots: Lot # 042408D08

Reason for recall

Fiber optic cable is missing from the Frederick's Converse Retractor Lot # 042408D08.

FDA-determined cause: Nonconforming Material/Component

Action taken

Consignee was notified by certified mail via an Urgent Medical Device Recall Notice dated 10/24/08, with return receipt. The letter identifies the product and states the issue with the product. The letter required the customer to discontinue use of the item and return it to Biomet Microfixation so that they may rework the instrument. It also asked the customer to notify any hospital personnel responsible for receiving recall notices of this action if the product had been further distributed. The RGA number should be utilized to return the products. The Inventory Reconciliation sheet should be filled out and faxed to 904-741-9425.

Details

Recall number
Z-0415-2009
Initiated
October 24, 2008
Posted by FDA
November 28, 2008
Terminated
September 11, 2009
Product code
FDG – Retractor, Fiberoptic
Quantity
3
Distribution
Nationwide Distribution --- including states of Chicago, IL and La Crosse, WI.
Recalling firm
Biomet Microfixation, Inc., Jacksonville, FL