Recall Z-0415-2024
Aizu Olympus Co., Ltd. · initiated October 16, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE
Codes / lots: UDI-DI 04953170305115 Serial Numbers 2635812, 2416423
Reason for recall
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
FDA-determined cause: Process control
Action taken
On October 16, 2023, the firm notified affected customers through URGENT MEDICAL DEVICE REMOVAL letters. Customers were instructed to inspect their inventory and identify affected devices. Affected devices should be quarantined and not used. Olympus Customer Solutions will contact customers by phone and email to arrange the return and repair of the affected product. If you require additional information, please do not hesitate to contact Olympus at 647-999-3203 or Cynthia.Ow@Olympus.com.
Details
- Recall number
- Z-0415-2024
- Initiated
- October 16, 2023
- Posted by FDA
- November 29, 2023
- Product code
- FDF – Colonoscope And Accessories, Flexible/Rigid
- Quantity
- 24 incorrectly repaired units
- Distribution
- Domestic distribution to FL, KY, MA, NJ, NY, and PA
- Recalling firm
- Aizu Olympus Co., Ltd., Aizuwakamatsu, 3 Chome
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0416-2024 Class II | GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… | 10/16/2023Open, Classified |
| Z-0417-2024 Class II | GIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… | 10/16/2023Open, Classified |
| Z-0418-2024 Class II | PCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… | 10/16/2023Open, Classified |
| Z-0419-2024 Class II | PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… | 10/16/2023Open, Classified |