Recall Z-0415-2024

Aizu Olympus Co., Ltd. · initiated October 16, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE

Codes / lots: UDI-DI 04953170305115 Serial Numbers 2635812, 2416423

Reason for recall

Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.

FDA-determined cause: Process control

Action taken

On October 16, 2023, the firm notified affected customers through URGENT MEDICAL DEVICE REMOVAL letters. Customers were instructed to inspect their inventory and identify affected devices. Affected devices should be quarantined and not used. Olympus Customer Solutions will contact customers by phone and email to arrange the return and repair of the affected product. If you require additional information, please do not hesitate to contact Olympus at 647-999-3203 or Cynthia.Ow@Olympus.com.

Details

Recall number
Z-0415-2024
Initiated
October 16, 2023
Posted by FDA
November 29, 2023
Product code
FDF – Colonoscope And Accessories, Flexible/Rigid
Quantity
24 incorrectly repaired units
Distribution
Domestic distribution to FL, KY, MA, NJ, NY, and PA
Recalling firm
Aizu Olympus Co., Ltd., Aizuwakamatsu, 3 Chome

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0416-2024 Class IIGIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… 10/16/2023Open, Classified
Z-0417-2024 Class IIGIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… 10/16/2023Open, Classified
Z-0418-2024 Class IIPCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… 10/16/2023Open, Classified
Z-0419-2024 Class IIPCF-H190L, EVIS EXERA III COLONOVIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the… 10/16/2023Open, Classified