Recall Z-0418-2020

Cincinnati Sub-Zero Products LLC, a Gentherm Company · initiated July 2, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Blanketrol II Hyper-Hypothermia System, Models 222R and 222S - Product Usage: The system is composed of a heater, a compressor, a circulating pump and blankets/pads.

Codes / lots: Model 222R (071-18490 through 123-20534) Model 222S (104-2-00001 through 161-2-01848)

Reason for recall

Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes

FDA-determined cause: Labeling False and Misleading

Action taken

Gentherm issued URGENT Medical Device Correction, Blanketrol II letters were mailed by USPS starting July 25th and Blanketrol III letters were mailed by USPS starting July 30th. Letters stated problem, health risk and action to take: INSTRUCTIONS TO CUSTOMERS: 1) Access updated manuals and ensure that obsolete manuals are removed from service. Updated manuals may be accessed via www.gentherm.com or physical copies may be requested from Gentherm Medical, LLC at (513)326-5320. 2) Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice. 3) Please complete and return the enclosed response form as soon as possible to acknowledge receipt of this notification and to inform Gentherm Medical, LLC that you have performed and completed the requested actions. Return the form by e-mail to FA2019-004/010@gentherm.com, or mail to: Gentherm Medical, LLC, 12011 Mosteller Road, Cincinnati, OH 45241. Questions, call Gentherm Medical Technical Support at 1-888-437-5608 or Stephanie Vocke @ (513)719-3262

Details

Recall number
Z-0418-2020
Initiated
July 2, 2019
Terminated
June 22, 2020
Product code
DWJ – System, Thermal Regulating
Quantity
3726
Distribution
Worldwide distribution - US Nationwide and countries of Afghanistan, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bhutan, Brazil, Brunei, Canada Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, France, Germany, Guatemala, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Nepal, Netherlands…
Recalling firm
Cincinnati Sub-Zero Products LLC, a Gentherm Company, Cincinnati, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0417-2020 Class IIBlanketrol III Hyper-Hypothermia System- body hypothermia system temperature management…Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage.… 07/02/2019Terminated
Z-0419-2020 Class IICoolBlue Hyper-Hypothermia System- body hypothermia system Model: Innercool - Product…Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage.… 07/02/2019Terminated