Recall Z-0425-2019

AROA Biosurgery, LTD. · initiated April 6, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

OviTex 1S Reinforced BioScaffold 16x20cm, Part Number F10256-1620G

Codes / lots: ERT-6H14 ERT-6J11 ERT-7K09 ERT-7L13

Reason for recall

Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.

FDA-determined cause: Other

Action taken

Beginning in April 2018, customers were visited by representatives and the affected units were replaced.

Details

Recall number
Z-0425-2019
Initiated
April 6, 2018
Terminated
April 15, 2019
Product code
FTM – Mesh, Surgical
Quantity
881 total
Distribution
The products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.
Recalling firm
AROA Biosurgery, LTD., Lower Hutt

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0415-2019 Class IIOviTex Reinforced BioScaffold 4x8cm, Part Number F10254-0408GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0416-2019 Class IIOviTex Reinforced BioScaffold 6x10cm, Part Number F10254-0610GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0417-2019 Class IIOviTex Reinforced BioScaffold 10x12cm, Part Number F10254-1012GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0418-2019 Class IIOviTex Reinforced BioScaffold 10x20cm, Part Number F10254-1020GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0419-2019 Class IIOviTex Reinforced BioScaffold 16x20cm, Part Number F10254-1620GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0420-2019 Class IIOviTex Reinforced BioScaffold 20x20cm, Part Number F10254-2020GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0421-2019 Class IIOviTex 1S Reinforced BioScaffold 4x8cm, Part Number F10256-0408GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0422-2019 Class IIOviTex 1S Reinforced BioScaffold 6x10cm, Part Number F10256-0610GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0423-2019 Class IIOviTex 1S Reinforced BioScaffold 10x12cm, Part Number F10256-1012GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0424-2019 Class IIOviTex 1S Reinforced BioScaffold 10x20cm, Part Number F10256-1020GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0426-2019 Class IIOviTex 1S Reinforced BioScaffold 18x22cm, Part Number F10256-1822GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0427-2019 Class IIOviTex 1S Reinforced BioScaffold 20x20cm, Part Number F10256-2020GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0428-2019 Class IIOviTex 2S Reinforced BioScaffold 6x10cm, Part Number F10258-0610GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0429-2019 Class IIOviTex 2S Reinforced BioScaffold 10x12cm, Part Number F10258-1012GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0430-2019 Class IIOviTex 2S Reinforced BioScaffold 10x20cm, Part Number F10258-1020GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0431-2019 Class IIOviTex 2S Reinforced BioScaffold 16x20cm, Part Number F10258-1620GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0432-2019 Class IIOviTex 2S Reinforced BioScaffold 18x22cm, Part Number F10258-1822GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0433-2019 Class IIOviTex 2S Reinforced BioScaffold 20x20cm, Part Number F10258-2020GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0434-2019 Class IIOviTex Reinforced BioScaffold 18x22cm, Part Number F10254-1822GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated
Z-0435-2019 Class IIOviTex 2S Reinforced BioScaffold 10x12cm, Part Number F10258-1012GDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed… 04/06/2018Terminated