Recall Z-0429-2022
bioMerieux, Inc. · initiated September 22, 2021
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
VIDAS Mumps IgG (MPG), REF 30218
Codes / lots: UDI 03573026064525, Lot #: 1008769120
Reason for recall
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.
FDA-determined cause: Nonconforming Material/Component
Action taken
For products with No remaining shelf life (taking into account the revised expiration date): - Identify impacted lots - Stop using the listed impacted lots - Destroy the remaining impacted lots in your inventory For products with remaining shelf life (taking into account the revised expiration date): - Identify impacted products and lots listed - Update product expiration date per your internal procedures of the remaining usable lots to meet the new revised expiration date specified in Table 1 - Continue product use until the revised expiration date. Consignees are asked to distribute this information to all appropriate personnel in their laboratory, retain a copy in their files, and forward this information to all parties that may use this product, including others to whom you may have transferred affected product. Complete the Acknowledgement Form in Attachment A and return it to USFieldActions@biomerieux.com or Fax to: (919) 620-3306 to confirm receipt of this notice.
Details
- Recall number
- Z-0429-2022
- Initiated
- September 22, 2021
- Product code
- LJY – Enzyme Linked Immunoabsorbent Assay, Mumps Virus
- Quantity
- One lot
- Distribution
- US Nationwide distribution.
- Recalling firm
- bioMerieux, Inc., Durham, NC
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0424-2022 Class II | VIDAS PRG Progesterone, REF 30409-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0425-2022 Class II | VIDAS Clostridium Difficile Toxin A&B, CDAB,REF 30118-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0426-2022 Class II | VIDAS Clostridium Difficile GDH, REF 30125-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0427-2022 Class II | VIDAS DEX@, Dimer Exclusion II, REF 30455-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0428-2022 Class II | VIDAS B.R.A.H.M.S. PROCALCITONIN PCT, REF 30450-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0430-2022 Class II | VIDAS Measles IgG (MSG), REF 30219bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0431-2022 Class II | VIDAS TOXO IgG Avidity (TXGA), REF 30222-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0432-2022 Class II | VIDAS RUB IgG (RBG), REF 30226bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0433-2022 Class II | VIDAS LH, REF 30406-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0434-2022 Class II | VIDAS FSH, REF 30407-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0435-2022 Class II | VIDAS Estradiol II, REF 30431-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0436-2022 Class II | VIDAS Lyme IgM II (LYM), REF 416436bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0437-2022 Class II | VIDAS Lyme IgG II (LYM), REF 417401bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0438-2022 Class II | VIDAS SARS-COV-2 IgM, Ref 423833-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |
| Z-0439-2022 Class II | VIDAS SARS-COV-2 IgG, Ref 423834-01bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It… | 09/22/2021Open, Classified |