Recall Z-0445-2020

Cook Inc. · initiated October 21, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiographic techniques.

Codes / lots: REF (RPN): HNB5.0-35-80-P-2S-C2 REF (GPN): G06622 Lot Numbers: NS9913495, NS9930288 UDIs: (01)00827002066229(17)220805(10)NS9913495, (01)00827002066229(17)220812(10)NS9930288

Reason for recall

The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.

FDA-determined cause: Process change control

Action taken

On October 21, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail instructing them to: 1. Examine inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: " Visit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to fax or mail. " Call the FDA at 800.FDA.1088. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your immediate assistance in this matter. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. We look forward to your response.

Details

Recall number
Z-0445-2020
Initiated
October 21, 2019
Terminated
June 30, 2020
Product code
DQO – Catheter, Intravascular, Diagnostic
Quantity
10
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX, NL, PK, PL, ZA, ES, SE, CH, TW, and GB.
Recalling firm
Cook Inc., Bloomington, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0421-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0422-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0423-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0424-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0425-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0426-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0427-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0428-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0429-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0430-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0431-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0432-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0433-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0434-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0435-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0436-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0437-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0438-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0439-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0440-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0441-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0442-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0443-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0444-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0446-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0447-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0448-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0449-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0450-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated
Z-0451-2020 Class IITorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and…The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches… 10/21/2019Terminated