Recall Z-0451-2014

Abbott Point Of Care Inc. · initiated October 28, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothrombin time test is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or Warfarin. The cartridge is to be used with the i-Stat 1 Analyzer and as part of the i-Stat System, the PT/INR test is to be used by trained health care professionals in accordance with a facility's policies and procedures.

Codes / lots: K020355 List number 04J50-01, 04J50-02; 03P89-24 Lot number R13149A; R13151; R13151A and cartridge lots from C13139 up to and including C13270A

Reason for recall

Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.

FDA-determined cause: Nonconforming Material/Component

Action taken

Abbott Point of Care Inc. (APOC) issued an Urgent Recall Notice on October 28, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of cartridges from PT/INR lots in this letter. All unused cartridges would be credited and are to be returned as instructed on the enclosed Business Reply Card. If customers provided cartridges from the lot number listed above to another institution they were instructed to provide a copy of this letter to them. Customers with questions were instructerd to contact Abbott Point of Care Technical at 1-800-366-8020, Option 1. For questions regarding this recall call 800-366-8020.

Details

Recall number
Z-0451-2014
Initiated
October 28, 2013
Posted by FDA
December 4, 2013
Terminated
May 29, 2014
Product code
GJS – Test, Time, Prothrombin
Quantity
794,928 units US; 119,472 units OUS
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Brazil, Chile, China, Finland, France, Germany, Greece, Hong Kong, Ireland, Italy, South Korea, Kuwait, Mexico, Netherland, New Zealand, Nigeria, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, south Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE, and United Kingdom,
Recalling firm
Abbott Point Of Care Inc., Princeton, NJ

Official FDA recall record