Recall Z-0452-2011

Integra LifeSciences Corp. · initiated September 29, 2010

Terminated

Product

DuraGen XS Dural Regeneration Matrix: 4 inches x 5 inches. Non Pyrogenic; Rx only;Integra LifeSciences Corporation, Plainsboro, NJ 08536 USA DuraGen XS is indicated as a dural substitute for repair of dura matter.

Codes / lots: Catalog number DXM1045, Lot number 1084359, 1091578, 1093431, 1094405, 1095442, 1095784, 1101583, 1101989.

Reason for recall

Test results no longer support a 3 year shelf life.

FDA-determined cause: Other

Action taken

Integra LifeSciences Corporation sent an URGENT Product Recall Notification letter dated September 29, 2010, to all affected customers. Integra NeruoSpecialists were informed of the recall via telephone conference call and recall notification letters sent by e-mail. Letters were also sent to customers by Federal Express with copies of e-mail provided to Interga Neuro Specialists on October 1, 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately quarantine any affected product and contact Integra Customer Service at 1-800-654-2873, select option 5, followed by option 3. Customers were instructed to fill out and return the enclosed "Recall Acknowledgement and Return Form" indicating whether or not they possessed any of the affected product. For any questions related to this recall call 1-800-654-2873 and selection option 1.

Details

Recall number
Z-0452-2011
Initiated
September 29, 2010
Posted by FDA
November 24, 2010
Terminated
April 23, 2012
Product code
GXQ – Dura Substitute
Quantity
571
Distribution
Nationwide Distribution including AK, AL, AZ, CA, CO, DC, DE, FL, GA, HI, IL, IN, LA, MA, MD, ME, MN, MO, MT, NC, ND, NH, NJ, NM, NY, OK, PA, TN, TX, UT, VA, VT, WA, WI, AND WV.
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0450-2011 DuraGen XS Dural Regeneration Matrix: 2 inches x 2 inches. Non Pyrogenic; Rx…Test results no longer support a 3 year shelf life. 09/29/2010Terminated
Z-0451-2011 DuraGen XS Dural Regeneration Matrix: 3 inches x 3 inches. Non Pyrogenic; Rx…Test results no longer support a 3 year shelf life. 09/29/2010Terminated
Z-0453-2011 DuraGen XS Dural Regeneration Matrix: 5 inches x 7 inches. Non Pyrogenic; Rx…Test results no longer support a 3 year shelf life. 09/29/2010Terminated
Z-0454-2011 DuraGen XS Dural Regeneration Matrix: 1 inch x 3 inches. Non Pyrogenic; Rx only;Integra…Test results no longer support a 3 year shelf life. 09/29/2010Terminated