Recall Z-0452-2016

Zimmer, Inc. · initiated November 10, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Zimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Double Trigger Handpiece. Rx only Made in Switzerland

Codes / lots: Model # 01-8507-400-00 & Model # 89-8507-400-00

Reason for recall

The firm identified a malfunction which made it possible for a handpiece to start by itself when the power source is connected.

FDA-determined cause: Device Design

Action taken

Zimmer sent an Urgent Medical Device Correction letter dated November 9, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Review this notification. 2. You will receive copies of notification sent directly to hospital risk managers and surgeons in your territory. As necessary, review and facilitate understanding of this notification by those entities. 3. Further identify hospital risk managers and/or surgeons in your territory who should also receive notification of this update, and supply the information to the entities you have identified. 4. If after reviewing this notification you have further questions or concerns, please call 330-364-0989 between the hours of 8 a.m. and 5 p.m. EST, Monday through Friday. Zimmer instructed their direct customers that they will upgrade the device with the new electronic during their annual preventative maintenance activities. Until the new upgrade is installed, the firm instructed their customers to follow the Instructions for Use.

Details

Recall number
Z-0452-2016
Initiated
November 10, 2015
Posted by FDA
December 21, 2015
Terminated
February 19, 2016
Product code
HAB – Saw, Powered, And Accessories
Quantity
87
Distribution
US Distribution to the states of : AR, CA, CO, CT, FL, GA, HI, IL, KS, KY, LA, ME, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, PA, RI, SC, TX, UT, VA, WA, WI, WV and WY.
Recalling firm
Zimmer, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0453-2016 Class IIZimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Single Trigger…The firm identified a malfunction which made it possible for a handpiece to start by itself when the power… 11/10/2015Terminated