Recall Z-0457-2009

Radiometer America Inc · initiated September 5, 2008

Terminated

Product

ABL7XX Calibration 1 Solution S1720. P/N S1720, 944-024 Intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb, FHbF). Also, intended for in vitro testing of samples of expired air for the parameters pO2 and pCO2.

Codes / lots: P/N S1720, 944-024; Lot: WA-04

Reason for recall

The firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the calibration curves for these parameters and thereby the measured results for patient samples and quality controls will be biased.

FDA-determined cause: Process control

Action taken

On, 9/2/2008 the firm sent out recall notification letters requesting their customers discard any Calibration 1 Solution (CAL1) S1720 (part number 944-024), lot WA-04, complete the attached return form and fax it to 800-736-0601. The firm will send out replacements for all affected product that the firm shows was shipped to their location. Further questions could be directed to Technical Support at 800-736-0600, option 2.

Details

Recall number
Z-0457-2009
Initiated
September 5, 2008
Posted by FDA
December 3, 2008
Terminated
April 25, 2012
Product code
JFP – Electrode, Ion Specific, Calcium
Quantity
3,200/200ml bottles
Distribution
Worldwide Distribution --- USA including states of AZ, CA, FL, IA, ID, IL, IN, KY, LA, MD, NC, ND, NE, NY, OH, OK, PA, SD, TX, VA, WA, and WI, and country of Canada.
Recalling firm
Radiometer America Inc, Westlake, OH

Official FDA recall record