Recall Z-0461-2011

Siemens Healthcare Diagnostics, Inc. · initiated April 12, 2010

Terminated

Product

The Dimension NTP and LNTP Flex reagent cartridge (N- Terminal Pro- Brain Natriuretic Peptide and N-Terminal Pro brain Natriuretic Peptide (low volume)

Codes / lots: Lot# FD0110 Exp# 04/20/2010, Lot#EB0122, EXP# 05/02/2010, Lot# FC0190, EXP# 07/09/2010, Lot# FC0219, EXP# 08/07/2010, Lot# FB0256, EXP#09/13/2010, Lot# FC0100, EXP#04/10/2010, Lot#EA0122, EXP#02/05/2010, Lot# FB0190, EXP# 07/09/2010, Lot# FB0219, EXP# 08/07/2010, lot# FA0256, EXP# 09/13/2010.

Reason for recall

Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.

FDA-determined cause: Other

Action taken

Siemens issued an "Urgent Field Safety Notice" dated April 2010, informing users of the problem and instructing them to inspect cartridges for leaking fluid. Cartridges with leaking fluid should be reported to the Siemens Technical Solutions Center. The Technical Solutions Center can be contacted at 800 441 9250.

Details

Recall number
Z-0461-2011
Initiated
April 12, 2010
Posted by FDA
November 24, 2010
Terminated
March 19, 2013
Product code
NBC – Test, Natriuretic Peptide
Quantity
3364
Distribution
Worldwide Distribution: USA, Europe, Australia, China, Japan, Korea, and Saudi Arabia.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Newark, DE

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0458-2011 The Dimension FT4L Flex reagent cartridge( Free Thyroxine) manufactured by Siemens…Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high… 04/12/2010Terminated
Z-0459-2011 The Dimension TSHL Flex reagent cartridge (Thyroid Stimulating Hormone) manufactured by…Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high… 04/12/2010Terminated
Z-0460-2011 The Dimension TNI Flex reagent cartridge (Cardiac Troponin), manufactured by Siemens…Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high… 04/12/2010Terminated