Recall Z-0466-2015

Remel Inc · initiated October 23, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Oxoid Legionella Latex Test, DR0800M, containing 50 tests per box. The box contains a variety of components, one of which is Oxoid L. pneumophila Latex Reagent, DR0802, Lot 1301912. The responsible firm on the label is Oxoid Ltd.

Codes / lots: DR0800M box lot 1316151, Exp. 30Sep2014

Reason for recall

A reagent within the test may return false negative results.

FDA-determined cause: Other

Action taken

ThermoFisher sent a Medical Device Recall letter dated October 23, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to notify any personnel who need to be aware of the potential for false negative results for the affected product. Customers should review their inventory for the affected products and discard them. Return the attached Medical Device Recall Return Response acknowledging receipt of the notice and disposal of the affected product. If product was further distributed customers should contact those entities, advise them of the situation and provide them with a copy of this letter. Customers with questions were instructed to contact their Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International).

Details

Recall number
Z-0466-2015
Initiated
October 23, 2014
Posted by FDA
December 16, 2014
Terminated
January 22, 2015
Product code
LIB – Device, General Purpose, Microbiology, Diagnostic
Quantity
59/50 test boxes
Distribution
Nationwide Distribution including AZ, CA, DC, FL, IL, ME, NJ, NY, PA, VA, and VT.
Recalling firm
Remel Inc, Lenexa, KS

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0465-2015 Class IIOxoid Legionella Pneumo Groups 2-14 Latex Test, DR0802M, containing 50 tests per box.…A reagent within the test may return false negative results. 10/23/2014Terminated