Recall Z-0468-2016
Siemens Healthcare Diagnostics, Inc. · initiated November 19, 2015
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualitative and Quantitative Factor Deficiency
Codes / lots: Lot Number 44637, exp. 2016-11-30
Reason for recall
Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Systems. This could lead, under certain circumstances, to an erroneously increased result with a Factor VIII Chromogenic Assay.
FDA-determined cause: Employee error
Action taken
Siemens sent an Urgent Medical Device Correction Letter dated November 19, 2015, to all affected customers . The letter identified the product the problem and the action needed to be taken by the customer. Actions to be Taken by Customers: If you are using the affected lot of Factor VIII Chromogenic Assay on a Sysmex CA-1500 or CA-7000 System, please discontinue use. - Due to a different detection principle, the use on BCS and BCS XP is not affected and you can continue using the product on BCS / BCS XP. - Complete and return the Field Correction Effectiveness Check attached to this letter within 30 days. - Review your inventory of this product to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the authorities. --- Corrective Action: The Factor VIII Chromogenic Assay Lot Number 44637 kits were placed on product hold on August 19, 2015, at the Siemens Distribution Centers. For further questions, please call ( 914) 524 - 2495.
Details
- Recall number
- Z-0468-2016
- Initiated
- November 19, 2015
- Posted by FDA
- December 22, 2015
- Terminated
- August 15, 2017
- Product code
- GGP – Test, Qualitative And Quantitative Factor Deficiency
- Quantity
- Domestic: 52 kits, Foreign: 623 kits
- Distribution
- Worldwide Distribution - US Distribution to the state of : Massachusetts and to the countries of : Austria, Czech Republic, France, Germany, Hungary, Italy, Poland, Serbia, Spain and Sweden.
- Recalling firm
- Siemens Healthcare Diagnostics, Inc., Tarrytown, NY