Recall Z-0468-2023

Zimmer, Inc. · initiated December 6, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

NexGen Option Stemmed Tibial Component Size 5, Item Number 00-5986-047-01

Codes / lots: All Unexpired Lots

Reason for recall

Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.

FDA-determined cause: Under Investigation by firm

Action taken

Urgent Medical Device Recall notification letters dated 12/6/22 were sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. It is recommended that for patients implanted with the NexGen Stemmed Option Tibial Component with either the LPS Flex or LPS Flex GSF Femoral Components, maintain an appropriate index of suspicion for patients with any new pain, inability to bear weight, swelling or instability of the knee. The UK NJR data suggests that tibial component loosening is a key cause of the increased TKR revisions. If your patient experiences any new pain or other symptoms related to the TKR, it is recommended that additional clinical and/or radiographic follow-up is completed. 3. If an impacted

Details

Recall number
Z-0468-2023
Initiated
December 6, 2022
Posted by FDA
December 9, 2022
Product code
JWH – Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Quantity
2460
Distribution
Worldwide - US Nationwide distribution.
Recalling firm
Zimmer, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0466-2023 Class IINexGen Option Stemmed Tibial Component Size 3, Item Number 00-5986-037-01Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial… 12/06/2022Open, Classified
Z-0467-2023 Class IINexGen Option Stemmed Tibial Component Size 4, Item Number 00-5986-037-02Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial… 12/06/2022Open, Classified
Z-0469-2023 Class IINexGen Option Stemmed Tibial Component Size 6, Item Number 00-5986-047-02Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial… 12/06/2022Open, Classified
Z-0470-2023 Class IINexGen Option Stemmed Tibial Component Size 7, Item Number 00-5986-057-01Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial… 12/06/2022Open, Classified