Recall Z-0474-2015

Abbott Ireland Diagnostics Division · initiated September 11, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 or 4 Bottle(s) (6.6 mL/27.0 mL) anti-T3 (sheep) coated microparticles in MES buffer with sheep IgG stabilizers. Minimum Concentration: 0.08% solids. Preservative: ProClin 300. - 1 or 4 Bottle(s) (5.9 mL/26.3 mL) T3 acridinium-labeled conjugate in citrate buffer with NaCl and Triton X-100 stabilizers. Minimum concentration: 0.33 ng/mL. Preservative: ProClin 300. The ARCHITECT Total T3 (TT3) assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of total triiodothyronine (Total T33) in human serum and plasma.

Codes / lots: Product List Numbers, Lot Numbers, and Expiration Dates: 1) List Number: 7K64-20, Lot Number 38901UI00, Expires: 29-Jan-15; 2) List Number: 7K64-25; Lot Number 38901UI01, Expires: 29-Jan-15

Reason for recall

17% of ARCHITECT Total T3 Reagent Kit, LN 7K64-20/-25 reagent lots 38901UI00 and 38901UI01 may exhibit lower Relative Light Units (RLUs) than expected, which in turn may result in controls out of range or patient results higher than expected.

FDA-determined cause: Employee error

Action taken

Abbott sent an Product Recall letter dated September 11, 2014, to all direct accounts (customers). The letters included instructions for customers to: 1) Discontinue use of and destroy any remaining inventory according to your laboratory procedures; 2) follow your laboratory protocol regarding the need for review of previously reported patient results; 3) replacement material is available; for ordering, please order through your normal ordering process; and, 4) if you have forwarded any of the affected lots to another laboratory, please provide them with a copy of this communication. Customers or healthcare providers with questions can contact Customer Service at 1-877-422-2688 (24 hours a day, 7 days a week).

Details

Recall number
Z-0474-2015
Initiated
September 11, 2014
Posted by FDA
December 3, 2014
Terminated
December 23, 2016
Product code
CDP – Radioimmunoassay, Total Triiodothyronine
Quantity
7,415 kits
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL , AR , AZ , CA , CT , FL , GA , IL , IN , KS , LA , MA , MD , MI , MN , MO , MS , MT , NC , NJ , NV , NY , OH , OK , PA , SC , SD , TN , TX , UT , VA , WA , WI and WV., and the countries of : Andorra , Angola , Argentina , Armenia , Austria , Azerbaijan , Bangladesh , Barbados , Belarus , Belgium , Bosnia &…
Recalling firm
Abbott Ireland Diagnostics Division, Lisnamuck Co. Longford

Official FDA recall record