Recall Z-0476-2019

Ortho-Clinical Diagnostics · initiated October 8, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS Chemistry Products dLDL Reagent-GENs 26 & 27. In Vitro Diagnostic to quantitatively measure LDL Cholesterol (LDLC) Product Code: 6801728 UDI (10758750006793)

Codes / lots: Lot Numbers: 1517-26-6396 (expires 6-Nov-2018) 1517-26-6582 (expires 2-Feb-2019) 1517-26-6639 (expires 2-Feb-2019) 1517-27-6808 (expires 7-May-2019)

Reason for recall

VITROS Chemistry Products dLDL Reagent GENs 26 and 27, negatively biased results

FDA-determined cause: Under Investigation by firm

Action taken

Ortho Clinical Diagnostics issue the customer notification on October 8, 2018, provides information regarding the potential for biased results when using VITROS dLDL Reagent, Generations (GENs) 26 and 27. Actions to take: For GEN 26: Discontinue using GEN 26 and switch to an alternate GEN of product as soon as possible. Please specify full or partial sales units remaining and indicate credit or replacement on your Confirmation of Receipt form. " For GEN 27: Install ADD DRV 6003 on your VITROS System upon receipt. Following a successful recalibration, resume using VITROS dLDL Reagent, GEN 27 remaining in your inventory. " Complete the Confirmation of Receipt form and return by October 22, 2018. Questions contact our Ortho Care" Technical Solutions Center at 1-800-421-3311

Details

Recall number
Z-0476-2019
Initiated
October 8, 2018
Terminated
October 13, 2022
Product code
MRR – System, Test, Low Density, Lipoprotein
Quantity
5850 units
Distribution
Nationwide Foreign: Australia Belgium Bermuda, HM11 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Poland Portugal Russia Singapore Spain Sweden The Netherlands United Kingdom
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY

Official FDA recall record