Recall Z-0477-2019
The Binding Site Group, Ltd. · initiated July 25, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgM antiserum, six calibrators and two controls. The calibrators and controls are based on stabilized human serum. All components contain sodium azide as a preservative and are provided in ready-to-use liquid format.
Codes / lots: Lot: 407483-1 and 407483-2
Reason for recall
After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.
FDA-determined cause: Under Investigation by firm
Action taken
On 07/25/2018, the Urgent Field Safety Notice and Field Correction E-back Form were emailed to customers. Customers were advised to cease use of the affected product and to dispose of the kits locally. This notice states that a witness signature of kit disposal is required on the E-back form. This notice needs to be passed on to all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred. Customers with additional questions are encouraged to contact their local firm representative, or customer service via phone at +44(0) 1214569696, or email at Technical.support@bindingsite.co.uk. The E-back Form requests that each customer and witness sign and date confirming they have both received and have followed the advice issued in the Notification. It also asks for the number of product in stock and number of kits disposed.
Details
- Recall number
- Z-0477-2019
- Initiated
- July 25, 2018
- Posted by FDA
- November 17, 2018
- Terminated
- April 27, 2021
- Product code
- CFN – Method, Nephelometric, Immunoglobulins (G, A, M)
- Quantity
- 108
- Distribution
- U.S.: IL, WV, NE, SD, NY, IN, WA, MD, NJ, PA, NC, AR, FL, VA
- Recalling firm
- The Binding Site Group, Ltd., Birmingham