Recall Z-0478-2013

Verathon, Inc. · initiated October 25, 2012

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

The GlideScope Video Laryngoscope (GVL) is a video laryngoscope that incorporates a high-resolution color camera, an LED light source, a rechargeable lithium battery, and video output to a remote monitor. It consists of a video laryngoscope and a portable video monitor. The affected GVL and Part numbers are GVL 3 0574-0007, GVL 4 0574-0001, and GVL 5 0574-0030. The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures.

Codes / lots: Serial numbers range for GVL3: 0574-0007/ MD105000 to MD112387; Serial numbers range for GVL4: 0574-0001/ LG105000 to LG112758; Serial numbers range for GVL5: 0574-0030/ XL105000 to XL111798.

Reason for recall

Re-usable video laryngoscope blades of the GlideScope GVL3, GVL4, and GVL5 were recalled due to potential cracking.

FDA-determined cause: Device Design

Action taken

The firm, Verathon, sent an "URGENT - Medical Device Recall" letter dated October 25, 2012, to their consignees/cutomers. The letter described the product, problem, and actions to be taken. The customers were instructed to return the recalled blades back to Verathon and complete and return the Device Recall Notification Reply Form via fax to (425) 883-2896 or email: cservice@verathon.com. All returned products will be quarantined and will be destroyed later. Verathon will replace the affected video laryngoscope blade at no cost to customers. Customers in Canada and US with questions can call 800-331-2313. If you are located outside of the US or Canada, please call 425-867-1348.

Details

Recall number
Z-0478-2013
Initiated
October 25, 2012
Posted by FDA
December 7, 2012
Terminated
September 27, 2013
Product code
CCW – Laryngoscope, Rigid
Quantity
total 2710 units
Distribution
Worldwide distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NV, NJ, NY, OK, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY; and countries of: Argentina, Australia, Canada, China, France, Germany, Indonesia, Japan, The Netherlands, Pakistan, Panama, Peru, Philippines…
Recalling firm
Verathon, Inc., Bothell, WA

Official FDA recall record