Recall Z-0478-2019

Voluntis · initiated October 8, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Insulia Diabetes Management Companion

Codes / lots: Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40

Reason for recall

The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.

FDA-determined cause: Software in the Use Environment

Action taken

On October 15, 2018, patients were notified that they could download the new version of the application from the Google Play Store. The notification was made through an automated message displayed on the application screen of all impacted patients requesting them to update their Insulia application. Patients were informed that the new version of the app resolved a an issue with the basal calculator identified on the Android version of the application. Patients who do not upgrade to the new version by November 9, 2018 will be blocked from using the app until they upgrade. For technical assistance or if you have any questions, please contact the Insulia support hotline.

Details

Recall number
Z-0478-2019
Initiated
October 8, 2018
Terminated
March 12, 2019
Product code
NDC – Calculator, Drug Dose
Quantity
14
Distribution
Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).
Recalling firm
Voluntis, Suresnes

Official FDA recall record