Recall Z-0494-2015

Aircraft Medical Limited · initiated November 4, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use, sterile video laryngoscope blade for use with McGRATH MAC Video Laryngoscopes

Codes / lots: Product Code: 350-017-000. Lot Numbers and Expiration Dates: Lot 13121202, Expires 2016/12/11; Lot 13122301, Expires 2016/12/25; Lot 13042401, Expires 2017/5/5; Lot 14030501, Expires 2017/3/16; Lot 14031814, Expires 2017/3/26; Lot 14061616, Expires 2017/6/25.

Reason for recall

The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming disconnected during intubation, specifically when the device is being retracted from the patients mouth.

FDA-determined cause: Component change control

Action taken

Aircraft Medical Limited sent a recall notification letter dated November 4, 2014 to the sole US consignee. The letter included instructions for the consignee to: 1) check its own stock and its customers' stock for the recalled products; 2) arrange for any remaining stock to be removed from use. Customers with any questions can contact Aircraft Medical by e-mail at ac@aircraftmedical.com.

Details

Recall number
Z-0494-2015
Initiated
November 4, 2014
Posted by FDA
December 2, 2014
Terminated
November 17, 2017
Product code
CCW – Laryngoscope, Rigid
Quantity
40,900 blades (US); 16,000 blades (Foreign)
Distribution
Worldwide Distribution - US Nationwide in the state of GA and the countries of Austria, China, Denmark, Finland, France, Hong Kong, Israel, Italy, Netherlands, Poland, Saudi Arabia, Singapore, South Korea, Taiwan, and Turkey.
Recalling firm
Aircraft Medical Limited, Edinburgh