Recall Z-0501-2007

Abbott Laboratories MPG · initiated August 11, 2006

Terminated

Product

IMx Sirolimus Reagent Pack; list 5C91-21; 100 test pack containing 1 bottle (9.7 mL) Anti-Sirolimus Antibody Coated Microparticles, 1 bottle (9.7 mL) Sirolimus Alkaline Phosphatase Conjugate, and 1 bottle (10 mL) 4-Methylumbelliferyl Phosphate, 1.2 mM; Produced by Axis-Shield Diagnostics, Ltd., Dundee, UK for Abbott Diagnostics Division, Distributed by Abbott Laboratories, Abbott Park, IL 60064 USA

Codes / lots: list 5C91-21, lots 053389, 053457, 063650, 063973, 064459, 064772, 064750, 065076

Reason for recall

The sirolimus values may shift higher after storage at 2-8C or after one freeze/thaw cycle of specimens.

FDA-determined cause: Other

Action taken

Abbott initiated a field correction recall by letter dated 8/11/06 to all customers receiving the affected reagent. The accounts were informed of the shift in sample values after storage at 2-8*C or with freeze thaw of specimens with the affected lots of reagents. The accounts were requested to review the revised sample handling information included in the attachment to the letter with their laboratory personnel and retain a copy of the letter and attachment for their records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.

Details

Recall number
Z-0501-2007
Initiated
August 11, 2006
Posted by FDA
February 15, 2007
Terminated
January 18, 2008
Product code
NRP – Sirolimus Test System
Quantity
2,800 kits
Distribution
Worldwide, including USA, Puerto Rico, Poland, Croatia, Latvia, Estonia, Serbia & Montenegro, Japan, Turkey, Singapore, China, India, Taiwan, Thailand, South Africa, South Korea, Australia, New Zealand, Norway, Italy, Portugal, Spain, Greece, Panama, Hungary, Slovakia, Russia, Romania, Cyprus, Germany, Lithuania, Saudi Arabia, Jordan, Israel, Lebanon, Kuwait, Oman, Qatar, UAE, Egypt, Mexico…
Recalling firm
Abbott Laboratories MPG, Abbott Park, IL

Official FDA recall record