Recall Z-0512-2012

Teco Diagnostics · initiated October 14, 2011

Terminated

Product

Urine Reagent Strips (URS) 11 Parameters URS-11-100G-001 510(K) number: K970250 Devise Listing number: D072377 Each Device consists of 100 strips of Urine Reagent Strips 11 parameters packed with a desiccant pouch inside a black bottle with Cortez Diagnostics Label on the bottle and the bottle placed inside a Cortez Diagnostics Box.

Codes / lots: Product # 2: URS-II-I00G-00I i. Lot Number: 35224 ii. Expiration Date: 08/2013

Reason for recall

This recall was initiated in response to findings during a FDA inspection of the firm. The firm's recall has been initiated due to improper manufacture of sub-component nitrite paper Lot 35134, Expiration Date: 01/2014. Discrepancies were found on the information recorded on formulation sheet and manufacturing log sheets. Sampling size submitted for QC testing was insufficient based upon number o

FDA-determined cause: Nonconforming Material/Component

Action taken

Teco Diagnostics sent an "URGENT: MEDICAL DEVICE RECALL" letter dated October 14, 2011 to affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to examine their inventory for the affected product and to quarantine any product found. A Recall Return Response Form was included with the letter for customers to complete and return. Contact the firm's Technical Support Department at (800) 222-9880 for questions regarding this recall.

Details

Recall number
Z-0512-2012
Initiated
October 14, 2011
Posted by FDA
January 11, 2012
Terminated
April 9, 2013
Product code
KQO – Automated Urinalysis System
Distribution
Distributed in CA and Poland.
Recalling firm
Teco Diagnostics, Anaheim, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0511-2012 Urine Reagent Strips (URS) 10 Parameters URS-l0-100W 510(K) number: K970250 Devise…This recall was initiated in response to findings during a FDA inspection of the firm. The firm's recall has… 10/14/2011Terminated