Recall Z-0513-2022
Straumann USA LLC · initiated November 30, 2021
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Neodent Easypack Helix GM Acqua 3.5X13/Smart 4.5X2.5 Article Number: 138.107
Codes / lots: Lot GNV20: UDI:7899878052906
Reason for recall
Mislabeled: Carton box item 138.089- Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 has an implant with measure of 3.5x13mm; and Carton Box (138.107- Easypack Helix GM Acqua has a 3.5x8mm implant instead of a 3.5x 13mm implant
FDA-determined cause: Packaging change control
Action taken
Neodent issued Medical Device Recall letter initial contact by phone initiated on November 30th, 2021; followed by Confirmation written letter via UPS December 3rd 2021. Letter states reason for recall, health risk and action to take: Search in your stock for units according to the description on the table above. " If the implant has not yet been used, identify and segregate the affected units, open or closed, in your stock. Return the product using the label included in this mailing, refer to Field Action . Upon receipt of the product by Straumann, replacement product will be sent. If the product is not found in your inventory, indicate on the Form below. " If the product has been implanted and no issue was identified, it is not necessary to remove, additional patient follow-up is not required. Indicate this on the Customer Confirmation Form (included) " If the product has been installed and an issue has been reported by the patient, for example: numbness, pain or any other symptom, please communicate immediately to Straumann. " For either cases, complete and return the enclosed Customer Confirmation Form to Action 003/2021. 3.1. DATA FOR PRODUCT RETURN: Straumann USA Regulatory Affairs Action 003/2021 Product Return 60 Minuteman Road, Andover, MA 01810
Details
- Recall number
- Z-0513-2022
- Initiated
- November 30, 2021
- Terminated
- February 5, 2025
- Product code
- DZE – Implant, Endosseous, Root-Form
- Quantity
- 1 unit
- Distribution
- US Nationwide Distribution in the states of FL, IL, OH
- Recalling firm
- Straumann USA LLC, Andover, MA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0512-2022 Class II | Neodent Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 -intended for dental implant Item…Mislabeled: Carton box item 138.089- Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 has an implant with measure… | 11/30/2021Terminated |