Recall Z-0519-2012

Smith & Nephew, Inc. Endoscopy Division · initiated July 26, 2011

Terminated

Product

Smith & Nephew, SPIDER2 Limb Positioner, including accessories Battery, foot pedal and charger Catalog Number: 72203299. Table, Operating-room, Pneumatic.

Codes / lots: Serial Numbers: VAL07, VAL08,VAL11,VAL15, VAL16, VAL19, VAL20, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54, 55, 56, 58, 59, 60, 61, 62, 63, 66, 71, 73, 76, 77, 79, 80, 81, 82, 88, 90, 92, 93, 94, 95, 97, 98, 99, 100, 101, 102, 103, 104, 106, 107, 110

Reason for recall

SPIDER2 may unlock (loss of traction) resulting in potential movement of the patient when the device is within close proximity to an RF source. Loss of traction can potentially result in the unintended repositioning of the extremity and potential for tissue or cartilage damage

FDA-determined cause: Device Design

Action taken

Smith & Nephew Advanced Surgical Devices Division sent an "URGENT- PRODUCT RECALL NOTIFICATION-URGENT" letter dated July 26, 2011 to all affected customers. The letter described the product, problem, and the actions to be taken by the customers. Customers were instructed to complete and return a form via fax at 405-917-8650. Contact the firm at 405-917-8556 for questions regarding this recall.

Details

Recall number
Z-0519-2012
Initiated
July 26, 2011
Posted by FDA
January 11, 2012
Terminated
June 10, 2016
Product code
FQO – Table, Operating-Room, Ac-Powered
Quantity
52 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada and France.
Recalling firm
Smith & Nephew, Inc. Endoscopy Division, Andover, MA

Official FDA recall record