Recall Z-0522-2020

Zimmer Biomet, Inc. · initiated August 21, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

HENNESSY LEFT CONST GLENOID HA, Item No. PM555360 UDI (01)00880304819597(17)260425(10)664030 Product Usage: Constrained Shoulder.

Codes / lots: 664030

Reason for recall

Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.

FDA-determined cause: Environmental control

Action taken

On September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.

Details

Recall number
Z-0522-2020
Initiated
August 21, 2019
Terminated
July 21, 2021
Quantity
219988 units (total)
Distribution
US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
Recalling firm
Zimmer Biomet, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0519-2020 Class IICompress, Mini Compress; Item Nos. 178350 178351 178352 178353 178354 178355 178356…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0520-2020 Class IICOMP RVRS Shoulder Baseplates, including Custom Products, Item numbers 115331S 115331…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0521-2020 Class IIACROMIUM FLANGE SET (Custom product) Item No. PM555148 - Product Usage: Arcomium…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0523-2020 Class IIVarious Hip systems, Item Nos. 104207 104212 104215 11-162809 11-162810 11-162811…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0524-2020 Class IIComprehensive HHR Tess Cleat, Item Nos. 110260 110261 110262 110263 Product Usage: N/AElevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0525-2020 Class IIComprehensive RS Cleat CAP, Item No. 110300Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0526-2020 Class IIVarious Custom Products, Item Nos. CP102940 CP102941 CP102942 CP102943 CP102944 CP102945…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0527-2020 Class IICopeland HA Resurfacing Head, MB/HA EAS Head; Item Nos. 11-114641 11-114642 11-114643…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0528-2020 Class IIHA Pins (DFS HA Cort Screw), Hoffman/Colles Pins (Colles CSER Sterile w/ Pin Kit) Item…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0529-2020 Class IIGUIDE WIRE 2.2MM X 28IN, Aim/Art Guide wire, Item No. 809222028 - Product Usage: These…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0530-2020 Class IIBALL NOSE GUIDE WIRE 80CM, 100CM. Item Nos. 281001080 281001100 Product Usage: These…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0531-2020 Class II2.0MM BALL NOSE GUIDE WIRE, Item Nos. 281017006 - Product Usage: These instruments and…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0532-2020 Class IILT DISC ULNA - HA/PC; RT DISC ULNA - HA/PC; LT DISC HMRL- HA/PC; RT DISC HMRL- HA/PC…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated
Z-0533-2020 Class IIBIOMOD PORCOAT HA GLENTRAY; Custom Shoulder Products Item Nos. PM0000097 PM555140…Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. 08/21/2019Terminated