Recall Z-0528-2025
Smiths Medical ASD, Inc. · initiated October 28, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
Codes / lots: REF 14-504-68, UDI/DI 35021239002586, Lot Numbers: 0003393, 0003394, 0003357, 0003358, 0003267, 0003368, 0003634, 0003254, 0003369, 0003165, 0003184, 0003196, 0003204, 0003206, 0003226, 0003256, 0003300, 0003307, 0003308, 0003314, 0003322, 0003359, 0003360, 0003361, 0003362, 0003363, 0003364, 0003377, 0003380, 0003635, 0003643; REF 14-504-68JP, UDI/DI 35021239002586, Lot Numbers: 0003369, 0003377, 0003393, 0003394, 0003634, 0003635, 0003643
Reason for recall
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
FDA-determined cause: Under Investigation by firm
Action taken
Smiths Medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 10/28/2024, mailed using a traceable means. The notice explained the issue, potential risk, and requested the consignee locate and destroy all affected products, and distribute the notice to all affected parties. Distributors were directed to notify their customers. For questions: About the communication: smithsmedical7602@sedgwick.com, 1-800-913-3897; To report adverse events: globalcomplaints@icumed.com, 1-866-216-8806
Details
- Recall number
- Z-0528-2025
- Initiated
- October 28, 2024
- Posted by FDA
- November 25, 2024
- Product code
- BSR – Stylet, Tracheal Tube
- Quantity
- 10,223 units
- Distribution
- Worldwide distribution.
- Recalling firm
- Smiths Medical ASD, Inc., Minneapolis, MN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0527-2025 Class II | smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17…Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The… | 10/28/2024Open, Classified |
| Z-0529-2025 Class II | smiths medical PORTEX Tracheal Tube Exchange Guide, 10Ch x 70cm, REF 14-504-76Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The… | 10/28/2024Open, Classified |